Автоклав getidy 23l коды ошибок

APPENDIX 9:Alarm code and solutions of JN-23 Steam Sterilizer

Alarm Code

Alarm Cause

Abnormally Exit

Power outage

E 98

during operation

Teperature sensor

(steam generator)

E 01

is broken!

Teperature sensor

E 02

(heating coil) is

broken!

Teperature sensor

E 03

(inside chamber)

is broken!

Alarm Code

Release Alarm

Alarm Cause

If need abnormally exit during sterilization,you have to press

«

«

«

ok

,LCD appear

E00 alarm code and flashing.Press

Alarm,and start vacuuming-drying process.After 3 minute evacuating,the sterilizer will stop

working.This cycle is over,the operation display shows main interface.The door of sterilizer can

be opened.

If you encounters abrupt power failure during the sterilizing operation,when the power is

restored,

the alarm code

Code.

Measure:temperature sensor resistance value( ) 1000+(Ambient temperature

(1)if measurement result is OK,check poor contact point between the sensor and the

control board.(2)if measurement result is No,replacing the temperature sensor.

Measure:temperature sensor resistance value( ) 1000+(Ambient temperature

(1)if measurement result is OK,check poor contact point between the sensor and the

control board.(2)if measurement result is No,replacing the temperature sensor.

Measure:temperature sensor resistance value( ) 1000+(Ambient temperature

(1)if measurement result is OK,check poor contact point between the sensor and the

control board.(2)if measurement result is No,replacing the temperature sensor.

Page 55 of 58

Solution

«

will appear on the LCD. Press the «

Ω ≈

Ω ≈

Ω ≈

«

«

button,then click

«

«

button again to relieve

«

button to clear this

*3.8)

*3.8)

*3.8)

Если в процессе работы автоклава возникают ошибки, встроенная система блокирует часть функций, отключает режимы. В большинстве случаев при наличии ошибок нормального пользоваться стерилизатором невозможно. Более того — это может быть опасно не только для стерилизуемой загрузки, но и для оператора автоклава.

Ремонт и устранение ошибок автоклава — задача квалифицированного инженера, профильного сервисного центра. Для успешного восстановления функциональности и безопасности автоклава необходимо не только общее понимание принципов работы паровых стерилизаторов, но и знание особенностей конкретных моделей, типовых проблем и методов их решения.

Одна из наших специализаций — сложный ремонт стерилизационного оборудования.

Обращайтесь, мы оперативно решим все возможные проблемы с вашим оборудованием.

Частые ошибки автоклавов. Коды и описания ошибок паровых стерилизаторов


В данном материале собраны списки кодов ошибок различных автоклавов и описаны самые распространенные из них. 

Важно. Если вы видите на экране (или в распечатке встроенного принтера автоклава) ошибку — сработала система защиты и самодиагностики стерилизатора.


Система контроля ошибок работы автоклава. Самодиагностика паровых стерилизаторов.

В современные автоклавы устанавливается огромное количество датчиков для всех систем: от закрытия и блокировки двери во время стерилизации до качества используемой воды, температуры и давления на различных этапах стерилизационного цикла в рабочей камере, парогенераторе и других частях автоклава.

Встроенный контроллер на основной плате управления постоянно мониторит все эти параметры, а также — динамику (скорость) их изменения. Например, если температура или давление в камере для основного цикла стерилизации поднимаются или сбрасываются очень медленно, автоклав не перейдет к следующему этапу.

Как отображаются ошибки на экране автоклава

Автоклав выдает ошибку? Как правило, даже самые дорогие и многофункциональные премиум-стерилизаторы оснащаются не очень большим дисплеем (а модели среднего и начального уровня могут иметь только цифровое табло), поэтому общепринятая практика — отображать на дисплее численно-буквенный код ошибки работы автоклава. Часто — с буквенным обозначением E, Err, ERROR, например E5, E7, Er02, Err5 и т.д.

Расшифровку возможных ошибок можно найти в сервисных инструкциях или на сайтах производителей стерилизаторов. В данном материале мы также приводим официальные расшифровки самых “популярных” ошибок автоклавов. 

При этом важно понимать, что формулировка ошибки указывает инженеру только направление поиска неисправности (а в некоторых случаях истинная причина может скрываться в совершенно другом блоке, поэтому даже зная код ошибки и его точное значение, всегда необходимо проводить полную диагностику всего автоклава для выявления всех неисправностей.

Ошибки автоклава

Какие ошибки при работе автоклава появляются чаще всего

У каждого производителя и каждой модели автоклава есть своя специфика, но по статистике основные (распространенные) типы ошибок чаще всего следующие:

  • Ошибка качества воды: используется некачественная, не дистиллированная вода, неисправен датчик воды или внутренние поверхности трактов, фильтры содержат инородные элементы, биоматериал.
  • Ошибка блокировки двери автоклава: не закрыта дверь, не работает механизм электронной блокировки, не срабатывают концевые датчики и т.д.
  • Ошибки температуры парогенератора и рабочей камеры автоклава: слишком высокая температура парогенератора, камеры, слишком быстро / медленно падает или поднимается температура. Причин этой группы ошибок может быть множество: от неисправностей самих датчиков но проблем с терморегуляторами и нагревательными элементами / ТЭНами.
  • Утечка вакуума, давления: отдельный и очень распространенный вид ошибок. Чаще всего связан с проблемами герметичности двери, уплотнителем, работой компрессоров и т.д.
  • Ошибки давления в рабочей камере автоклава: автоклав не держит давление, давление не поднимается, давление в камере сбрасывается слишком медленно и т.д.
  • Ошибки рабочих режимов автоклава: например, температура и давление не соответствуют друг другу, отсутствует предварительное вакуумирование, предварительный нагрев
  • Ошибки, прямо указывающие на неисправность автоклава: Например, сломан микровыключатель двери (концевой датчик), загрязнен фильтр и т.д.

Не важно, какую ошибку выдает автоклав, помните то, о чем мы писали выше: даже при явном указании на неисправность необходима полная проверка всего аппарата.

Коды ошибок популярных паровых стерилизаторов

Далее приводим таблицы — списки ошибок популярных автоклавов. Данная информация поможет сориентироваться в направлении поиска неисправностей, если вы видите одну из таких ошибок на дисплее своего автоклава — сообщите ее сервисному инженеру, это поможет в подготовке к ремонту / выезду.

Ошибки автоклавов Woson (Восон)

Очень часто к нам обращаются за помощью по ошибкам автоклавов Woson (например, TanzoC23). Ниже приводим список кодов ошибок на основе этой модели. Для многих автоклавов производителя описание схожее, но рекомендуем каждый раз сверяться с соответствующим руководством. 

Вы также можете написать нам на Email, Whatsapp, позвонить по номерам в контактах.
 

E01

Медленный нагрев

E02

Ошибка сброса давления

E03

Открытая дверь

E04

Низкое давление

E05

Медленный спад давления

E06

Медленное вакуумирование

E07

Окружающее давление

E08, E09

Микропереключатели передней и задней двери сломаны

E10

Низкий предварительный нагрев

E11

Высокая температура

E12

Низкая температура

E13

Ошибка автоматической подачи воды

E17

Ошибка качества воды

E23

Давление и температура не совпадают

E24

Вока не стерильна

E25

Ошибка воздушного фильтра

E27

Серьезная утечка

E31, E32, E33 E34

Ошибка термодатчиков 1 — 4

E51, E52, E53, E54

Датчик температуры высокий уровень выхода

E71

Сенсор давления низкий уровень выхода

E72

Сенсор давления высокий уровень выхода

E14

Ошибка несоответствия внутренних сенсоров / Повреждение сигнала

E99

Автоклав не набирает температуру в 134 градуса. / Экстренный выход (зависит от версии микропрограммы и других особенностей)

Проконсультироваться с инженером

Еще один вариант индикации ошибок автоклава Woson: с обозначениями Err / Er. Характерны для многих моделей.

Err7

Переработка по времени / Сверхурочная работа  — одна из самых распространенных ошибок, возникающих при работе автоклавов Woson. Необходима последовательная проверка герметичности камеры, клапанов, всех датчиков, помп и компрессоров, полная диагностика автоклава.

Err12

Ошибка вакуума (вакуумирования / вакуумной системы)

Er98

Электропитание автоклава было отключено во время выполнения рабочего цикла. Необходимо проверить силовые и управляющие цепи, уточнить, не было ли внеплановых / веерных отключений, колебаний парамертов электросети.

Er99 / Er00

Принудительный выход. Нормальный ход выполнения цикла был прерван оператором

Ошибки автоклава Woson Tanzo-E

Альтернативный вариант таблицы с кодами ошибок автоклавов Woson Tanzo-E

Er01

Превышены допустимые температуры парогенератора

Er02

Превышен температурный режим нагревательного кольца

Er03

Температура рабочей камеры за пределами нормы

Er04

Возникла ошибка в настройках параметров температуры и давления

Er05

Ошибка сброса давления (повышенное)

Er06

Ошибка двери рабочей камеры (открыта во время рабочего цикла)

Er07

 Аналогично параметру Err07. Переработка по времени

Er08

 Высокое давление

Er09

 Датчик температуры рабочей камеры фиксирует температуру за пределами нормы (низкая или высокая). Ошибка характерна для сисетм с двойными сенасорами.

Er10

 Температурный режим не соответстует давлению

E12

 Вакуумирование выполнено с ошибкой

Er14

 Показатели двойных сенсоров температуры находятся в слишком широком диапазоне (не соответствуют друг другу)

Ошибки автоклава DGM

Ошибки автоклавов производства DGM рассматриваем на примере популярной модели DGM 130. Данный производитель обозначает сообщения о сбоях как Er

Er2

Давление пара в системе подачи низкое

Er3

Высокая температура рабочей камеры

Er4

Слишком высокое давление рабочей камеры

Er5

Ошибка вакуумного насоса

Er6

Контактор PP выключен, нет питания

Er7

Ошибка температурного датчика в рабочей камере

Er8

Ошибка датчика давления в рабочей камере

Er9

Ошибка датчика давления в рубашке

Er12

Нет пара на входе стерилизатора

Er14

Ошибка температурного датчика в емкости PP

Er15

Низкий уровень воды в емкости PP

Er16

Ошибка защиты от пара

Er31 — 38

Ошибки двери

Дополнительно, на стерилизаторах со встроенным парогенератором возможны следующие сообщения:

Er21

Отсутствует вода для парогенератора

E22

Ошибка насоса или датчика воды парогенератора

E23

Ошибка времени прогрева

Er24

Ошибка отключения нагревателей

Er25

Ошибка термореле генератора

Ошибки автоклавов MELAG (Мелаг)

Современные серии автоклавов Melag Euroklav, Vacuklav в зависимости от модели могут комплектоваться как простым двухстрочным дисплеем для отображения текста, так и полноценным цветным экраном с довольно хорошим разрешением.

Помимо непосредственно кодов ошибок на дисплее могут отображаться системные сообщения, рекомендации и предупреждения (то есть неисправность еще не зафиксирована, но продолжение работы невозможно до принятия соответствующих действий).

Предупреждения
 

Warning! Door open

Неплотно закрыта дверца

Warning! No water

Уровень воды при внутренней подаче ниже минимальной отметки

Warning! No feed water

Вода не достигла датчика воды (при внешней подаче)

Wastewater tank Full

Заполнена камера отработанной воды

Water quality Poor

Низкое качество воды, но возможен ручной старт

Water quality Bad

Неприемлемое, опасное качество воды, старт невозможен

Exchange Sterile filter

Требуется замена фильтров (возможны и другие причины этого сообщения)

Printer is not ready / Printer memory full

нет связи с принтером / заполнена память встроенного принтера

Please service

Необходимо сервисное обслуживание

Информация об ошибках автоклавов Melag

Malfunction 1: Vacuum system

Ошибка скорости набора вакуума

Malfunction 2: Steam generator

Длительный нагрев, ошибка парогенератора

Malfunction 4: Pressure release

Давление долго не нормализуется

Malfunction 8: Timebase

Ошибка таймера, системный сбой

Malfunction 9: Door open

Открыта дверь

Malfunction 10: Steamgen. Too hot

Ошибка температуры или алгоритма работы парогенератора

Malfunction 12 Door lock

Проблемы с открытием дверцы

Malfunction 14: Water supply

Ошибка подачи воды

Malfunction 18: Sens. Defect Nr

Ошибки датчиков температуры, давления и качества воды

Malfunction 21: Preheating

Ошибка предварительного прогрева

Malfunction 22: Overheating

Слишком сильный предварительный прогрев

Malfunction 23: Current

Время нормализации давления больше максимального

Malfunction 26: A/D-Converting

Ошибка контроллегра / датчиков, платы управления

Malfunction 27: Temp. sens. def 1,2

Ошибка разницы температур термодатчиков

Malfunction 29: Battery RAM

Проблема на плате управления

Malfunction 33: Pressure drop

Длительный набор давления

Malfunction 34: Sterilization TU

Температура стерилизации недостаточная

Malfunction 35: Sterilization TO

Температура стерилизации превышает максимальную

Malfunction 36: Sterilization PU

Слишком низкое давление

Malfunction 37: Sterilization PO

Слишком высокое давление

Malfunction 38: Sterilization TD

Ошибка вычисления температуры

Malfunction 41: Flow drying

Время нормализации давления при сушке больше максимального

Malfunction 42: Drying press.-pump

Время набора давления при сушке больше максимального

Malfunction 43: Drying Vac.-pump

Время создания вакуума больше максимального

Связаться с инженером

Ошибки автоклавов MOCOM (Моком)

Ошибки автоклавов производства Mocom рассмотрим на примере серии, Neutra Vacuum/ Частично эти расшифровки применимы и к Mocom Exacta

В автоклавах Mocom все нештатные ситуации обозначаются сообщениями трех категорий: ошибки (E), аварийные ситуации (A), опасные ситуации (H)
 

E 010

DOOR OPEN / Открыта дверь

E 020

DOOR UNLOCKED / Дверь не закрылась вовремя

E 021

DOOR LOCKED / Дверь не открылась вовремя

E 030

WATER MIN / Низкий уровень воды

E 031

EXHAUST MAX / Бак отработанной воды заполнен

E 041

FILLING PROBLEM / Слишком интенсивная подача воды в бак

E 900

TEST FAILED / Ошибка вакуума (утечка)

E 901

TEST FAILED / Ошибка вакуума (ожидание)

E 902

TEST FAILED / Ошибка вакуума (таймаут импульса)

E 999

MANUAL STOP / Цикл остановлен вручную

A 022

LOCKING PROBLEM / Ошибка микровыключателей сдвери (Выкл)

A 023

LOCKING PROBLEM / Ошибка микровыключателей сдвери (Вкл)

A 024

LOCKING PROBLEM / Ошибка микровыключателей сдвери (Вкл / Выкл)

A 032

LEVEL PROBLEM / Ошибка датчика уровня

A 040

FILLING PROBLEM / Невозможна автоматическая заправка

A 101

PTC BROKEN / Повреждение в рабочей камере

A 102

PTC BROKEN / Повреждение в парогенераторе

A 111

PTC SHORTCIRCUIT / Короткое замыкание в рабочей камере

A 112

PTC SHORTCIRCUIT / Короткое замыкание в парогенераторе

A200

HEATING PROBLEM / Ошибка предварительного нагрева

A 250

PV1 TIMEOUT / 1 вакуумный импульс

A 251

ATM1 TIMEOUT / 1 поднятие давления до атмосферного

A 258

PPP TIMEOUT / 3 пульсовое колебание

A 259

PROCESS TIMEOUT / Ошибка фазы стерилизации 

A 260

PPD TIME-OUT / Сброс давления слишком медленный

H 150

MPX BROKEN / Сломан датчик давления

H 992

OVERHEATING PT2 / Избыточный нагрев парогенератора

В таблице указаны не все коды ошибок. Если вы не нашли здесь свой код ошибки — свяжитесь с нами.

Ошибки автоклавов Runeys

Ошибки автоклавов китайского производителя Runeys разберем на примере модели TINHERO 16L/B. Индикация ошибок данного автоклава очень простая, указывает только на неисправный блок, далее нужна диагностика.

E1

Неполадки сенсора

E2

Очень высокое давление

E3

Температура камеры больше 145 градусов

E4

Нестабильная температура и давление

E5

Ошибка подачи пара

E6

Ошибка двери

E7

Температура парогенератора

E8

Предварительный нагрев

E9

Ошибка подачи пара

Ошибки автоклавов Youjoy 

Китайский производитель стерилизационного оборудования Yoyjoy также оснащает свои автоклавы системой самодиагностики. Она довольно простая и позволяет в общем виде определиться с направлением поиска. Ошибки рассматриваем на примере автоклава BES-12L-B-LED

Е1

Ошибка сенсора температуры

Е2

Повышенное давление

Е3

Повышенная температура

Е4

Нет постоянной температуры

Е5

Блокировка паровой трубы

Е6

Ошибка двери

Е7

Перегрев генератора пара

Е8

Избыточное время нагрева

Е9

Избыточное время выдержки

давления

Ошибки автоклавов Tuttnauer

Израильский производитель автоклавов Tuttnauer известен довольно высокой надежностью, и большим количеством встроенной электроники, систем защиты. Ниже приводится таблица ошибок настольного автоклава Tuttnauer 2540 NOVA-3. В других моделях система контроля и структура ошибок могут быть намного сложнее (особенно это касается серии Hospital, проходных автоклавов большого объема).

Low Temp.

Температура не соответствует параметрам стерилизации

Low Heat

Невозможно достичь температуры стерилизации в течение 60 минут

High Temp

Температура на 3°C (и более) выше температуры стерилизации при цикле, поврежден датчик температуры

Low Pres

Давление ниже соответствующего температуре стерилизации

High Pres

Слишком высокое давление

Low Vacuum

Недостаточный уровень вакуума на стадии предварительного вакуумирования (менее 15 кПа)

Man. Stop

Ручная остановка цикла

Power Dn

Незапланированное отключение питания во время цикла стерилизации

Add Water, Low Water

В резервуаре низкий уровень воды, отключение нагревателя по иным причинам

Door Unlock

Проблема с герметичностью двери

Ошибки автоклавов PT Medical

Ошибки паровых стерилизаторов производителя PT Medical рассмотрим на примерах автоклавов серии BTD: BTD23L-A, BTD17L-A.

Е1

Длительный набор давления

Е2

Давление выше допустимого

Е3

Открыта дверь

Е4

Сбой давления в автоклаве

Е5

Длительный сброс давления

Е6

Дверной переключатель открыт или неисправен

Е7

Давление близко к атмосферному

Е8

Ошибка замка блокировки двери

Е9

Ошибка двигателя двери

Е10

Длительный нагрев

Е11

Нарушение автоматической подачи воды

Е12

Короткое замыкание внутреннего датчика температуры

Е13

Короткое замыкание внешнего датчика температуры

Е14

Короткое замыкание датчика парогенератора

Е15

Обрыв в цепи внутреннего датчика температуры

Е16

Обрыв в цепи внешнего датчика температуры

Е17

Обрыв в цепи температурного датчика парогенератора

Е18

Температура с внутреннего термодатчика выше 150℃

Е19

Температура с внешнего термодатчика выше 170℃

Е20

Температура парогенератора выше 230℃

В некоторых моделях (например у PT Medical BTD 23 количество сообщений больше. На дисплее может отображаться ошибка 31.

Заказать консультацию Вызвать инженера

Если вы встретились с незнакомой ошибкой на своем автоклаве или устранить проблему самостоятельно не удается — свяжитесь с нами по телефону или электронной почте- поможем оперативно найти причину и восстановить функциональность стерилизатора.


Все автоклавы для ЦСО, хирургии, стоматологии и других областей в едином каталоге Cordismed с описанием и характеистиками.

Паровые стерилизаторы

Теги:

Getidy KD Series Instruction Manual

  • Contents

  • Table of Contents

  • Bookmarks

Quick Links

EURO

C E R T

1128

0197

English

CLASS B and S

Steam Sterilizer

Instruction Manual

JN-23

MODEL:

loading

Summary of Contents for Getidy KD Series

  • Page 1
    EURO C E R T 1128 0197 English CLASS B and S Steam Sterilizer Instruction Manual JN-23 MODEL:…
  • Page 2
    Contents 1.Summary &Suitalbe Using Range……………………08 2.Technical parameter………………………..08 3.Installation…………………………..12 JN-23 steam sterilizer function and operation instruction …………….16 JN-23 steam sterilization program introduction ………………..27 6.Mantenance…………………………..38 7.Scraping……………………………45 APPENDIX 1:Preparing the instruments for sterilization………………45 APPENDIX 2:Packing………………………….46 APPENDIX 3:Positioning the load……………………..47 APPENDIX 4:Description of tests………………………49 APPENDIX 5:Validating the cycles…………………….51 APPENDIX 6:Quality of process water……………………52 APPENDIX 7:JN-23 steam sterilizer hydraulic circuit……………….53…
  • Page 3
    To buyers: Thank you very much for your choice of using our Desktop Pressure steam sterilizer! Before you use steam sterilizer,please fill in “User Report”firstly. Had it,we can check your using file promptly,so we can service you more quickly and effective! Page 02 of 58…
  • Page 4
    Page 03 of 58…
  • Page 5
    Page 04 of 58…
  • Page 6
    ZHEJIANG GETIDY MEDICAL INSTRUMENT CO.,LTD. ADDRESS:Duancun, Dongnan Industrial Zone, Shuxi Street, Wuyi, Jinhua, Zhejiang321200, China. TEL:0086-579-87712106 87707499 DOWNLOAD Web: www.getidy.com E-MAIL:getidy@getidy.com john@getidy.com Page 05 of 58…
  • Page 7
    Page 06 of 58…
  • Page 8
    Page 07 of 58…
  • Page 9
    TEST CYCLES: 3 CLASS: B and S ± CAPACITY OF CLEAN-WATER TANK: 4.5L POWER SUPPLY VOLTAGE: AC 230V 10% CAPACITY OF USED-WATER TANK: 3.5L MAINS FREQUENCY:50~60Hz GROSS WEIGHT: 65Kg POWER OUTPUT: 1600W NET WEIGHT: 59Kg CHAMBER: 23L φ PACKING SIZE(LxWxH):730mmx580mmx560mm DIAMETER x DEPTH: 250mmx450mm ABSORBED CURRENT: 10A…
  • Page 10
    JN-23 overall space required Page 09 of 58…
  • Page 11
    Usable space 23 litres=10.60litres Model:JN-23 Date of Manufacture Serial Number Page 10 of 58…
  • Page 12
    Page 11 of 58…
  • Page 13
    Page 12 of 58…
  • Page 14
    Page 13 of 58…
  • Page 15
    -The two front feet of the sterilizer are raised by 2 to 3 cm, resulting in the front end point being higher than the rear end point. -Artificially open the door to the steps: (1)Open the small cover of the sterilizer door.(Fig.3.1-5) (2)Insert the hexagon spanner into the door shaft and rotated clockwise until the door is fully opened.(Fig.3.1-6) (3)Remove the operation panel placed in the chamber and connect the operation panel and the control panel.
  • Page 16
    -Bring the instrument tray and accessories, remove the poly bag. -After you put medical instrument into the tray,turn on the power (green switch indicator light).(Fig.3.1-9) -Press the power revert button and LCD screen appears the main interface.(Fig.3.1-9) -Select the sterilization program you want, then press start to sterilization. -For details, please refer to Chapter 4.
  • Page 17
    4.JN-23 steam sterilizer function and operation instruction 4.1 JN series steam sterilizer contains 1-time(solid/plastic) and 3-times(hollow/prion/cotton) vacuum programs. With lager color LCD screen for showing pressure, temperature, time, running state and error warning. Temperature precision: 0.1℃ ;Pressure precision: 0.001bar. 4.2 Operator interface introduction Turn on the power at the bottom right,click the power revert button located on the right side of the middle.
  • Page 18
    4.3 Warning icon introduction Press “program ”(Fig.4.3-1),to enter the program interface(Fig.4.3-2).Select the program you want and click to enter. 4.3.1 Door status Fig.4.3-1 Fig.4.3-2 Fig.4.3-3 Fig.4.3-4 A)when you turn on the power and the door was closeing,it is show Fig.4.3-3 interface warning.It promps you to open the door and close the door again.
  • Page 19
    σ C)when it show Fig.4.3-7 interface warning,means that distilled water was unqualified ( >80μms) Prompts you to drain unqualified water in the clean-water tank and add the qualified distilled water again.The warning Will be eliminated. Water supply requirements: ●conductivity ≤15μs/cm(Room temperature) ●bleach content≤2mg/L ●PH:5~7 ●Hardness:≤0.02mmol/L…
  • Page 20
    Click“SET UP 1”(Fig.4.4-1)into Fig.4.4-2 interface.Select the program you want and click to enter. Fig.4.4-1 Fig.4.4-2 Fig.4.4-3 Fig.4.4-4 4.4.1.Language: Click “LANGUAGE”(Fig.4.4-3) into Fig.4.4-4 interface.click “ ”,Select language you want,then click“ ” to save and exit. 4.4.2.Date and time Click “Date and Time”(Fig.4.4-5) into Fig.4.4-6 interface.Click“ ”,change number of “Year,Month,Date,Hour, Minute,second”,then click “…
  • Page 21
    4.4.5.Preference: Sleepmode/Delay Start/Power Saving Click “Preference”(Fig.4.4-11) into Fig.4.4-12 interface. Fig.4.4-9 Fig.4.4-10 Fig.4.4-11 Fig.4.4-12 4.4.5.1.Sleepmode If the sterilizer was not operated for a long time, power will be disconnected for safety and saving power automatically. When you want to revert back to working, sterilizer will get rid of sleep mod after pressing the Revert Switch. “…
  • Page 22
    4.4.5.3.Power Saving Click “Power Saving”(Fig.4.4-17) into Fig.4.4-18 interface.Press “ ”to select power saving mode you want,then click“ ” to save and exit. “NO”:After each sterilization cycle is completed, the power will not be turn off. “Single”:The power saving will be activated only for this operation. But not effective for next cycle. “ALL”:After each sterilization cycle is completed, the power will be turn off.
  • Page 23
    4.4.6.2.Pipeline Cleaning Click“ Pipeline Cleaning”(Fig.4.4-23) into Fig.4.4-24 interface.Press “ ” buttom will start,into Fig.4.4-25 interface. Fig.4.4-25 Fig.4.4-26 Fig.4.4-27 Fig.4.4-28 WARN:Please after fill with distilled water,closing the door,then press the “ ” button,Start cleaning. Recommend cleaned once every 3 months. ATTENTION:Cleaning work will begin only when the steam generator temperature reaches 200℃ below the “…
  • Page 24
    4.4.6.5.Auto-Water Supply Click“ Auto-Water Supply”into Fig.4.4-31 interface.Press “ ” to start(Fig.4.4-32).If you need to stop halfway, press “ ” to exit. WARN:Before starting, please provide enough distilled water and plug in the inlet pipe to the valve. 4.5“SET UP 2” introduction 4.5.
  • Page 25
    Click“ Label Printer”(Fig.4.5-4) into Fig.4.5-5 interface.then press “Bluetooth signal connection”Fig.4.5-6 into Fig.4.5-7 interface.Click “ ”bottom to start connection.“CONNECTING…”:During the connection;“SUCCESS”:connection succeeded. Back to Fig.4.5-6 interface.click “Label Printer”(Fig.4.5-8) into Fig.4.5-9 interface. Click on “ ”to change the number of labels you want,Click “ ”bottom to print.
  • Page 26
    4.5.3. USB Record If you want to check the sterilization reports, please press “USB Record”(Fig.4.5-10) enter into interface Fig.4.5-11. “ ”: select the different cycle reports from 1~128; “ ” : move the page up/down. You can insert U disk to download the sterilization reports . Download the current a record Download all the records Fig.4.5-13…
  • Page 27
    4.6“Abnormally Exit”function If you wish to stop the running program, you can click the stop(Fig.4.6-1) and select OK(Fig.4.6-2) . The LCD screen will show E00 alarm code and alarm cause.After you click on start (Fig.4.6-3),into automatically.(Fig.4.6-4) After about 3 minutes, the program will be finished. When the screen show the sterilization program initial interface, you can open the door and bring out the medical instrument inside chamber.
  • Page 28
    5.JN-23 steam sterilizer program introduction Click“program”(Fig.5-1) into the sterilization program interface(Fig. 5-2) . 5.1 Sterilization program list Program sterilization sterilization drying vacuum Name temperature time time times ℃ 04min 09min 1 times 1.Solid ℃ 2.Hollow 04min 15min 3 times Fig.5-1 Fig.5-2 ℃…
  • Page 29
    5.2 Each sterilization phase diagram Heating coil preheating stage Start sterilization stage Pre-vacuum stage Exhaust stage After sterilization ONE PRE-VACUUM Heating stage First times Drying stage STERILIZATION PROGRAM Exhaust stage Balance atmospheric PROCESS SCHEMATIC and print sterilization Heating stage Second times record stage TIME Page 28 of 58…
  • Page 30
    THREE PRE-VACUUM STERILIZATION PROGRAM PROCESS SCHEMATIC TIME Heating stage The third times Heating coil preheating stage The econd times pre-vacuum stage Heating stage Start sterilization stage First times p re-vacuum stage The Second times Heating stage Exhaust stage The econd times Exhaust stage The First times After sterilization…
  • Page 31
    5.3 Run sterilization program 5.3.1 THREE TIMES PRE-VACUUM PROGRAM(HOLLOW PRION COTTON) “ ” Click start botton of Fig.5.3-1 interface,the sterilization program will be started and enter into Fig.5.3-2 or Fig.5.3-3 interface. When temperature of heating coil(T3) rised and less than 100℃,Screen will show Fig.5.3-2 interface (the heating coil pre-heating stage). When temperature of heating coil(T3) rised and over than 100℃,Screen will show Fig.5.3-3 interface (the first times pre-vacuum stage).
  • Page 32
    Fig.5.3-5 Fig.5.3-6 Fig.5.3-7 Fig.5.3-8 When the vacuum value reaches -0.700bar value,the control system will open water pump-solenoid valve(Ev2) and steam generator start to inject the steam into chamber.The temperatures(T1 and T2) and pressure inside the chamber will be rised at the same time. 5.3.1.6.THE SECOND TIMES EXHAUST STAGE(Fig.5.3-9) When the pressure inside the chamber reaches 1.100bar,the control system will open the exhaust-solenoid valve(Ev1) for exhausting untill pressure downs to P=0.000bar.LCD shows “00:01:29”time countdown.
  • Page 33
    Fig.5.3-13 Fig.5.3-14 Fig.5.3-15 Fig.5.3-16 5.3.1.10.AFTER STERILIZATION EXHAUST STAGE(Fig.5.3-13) When the “sterilization time”countdown is over,the control system will open the exhaust-solenoid valve(Ev1) for exhausting untill pressure downs to P=0.000bar.LCD shows “00:01:29”time countdown. 5.3.1.11.DRYING STAGE(Fig.5.3-14) When the pressure inside the chamber downs to 0.000bar,the control system will open the vacuum pump-solenoid valve(Ev4) and start evacuating and drying.LCD shows “00:14:59”time countdown.
  • Page 34
    PLASTIC ) 5.3.2 ONE TIMES PRE-VACUUM PROGRAM(SOLID “ ” Click start botton of Fig.5.3-19 interface,the sterilization program starts running, shows Fig.5.3-20 or Fig.5.3-21 interface. When temperature of heating coil(T3) rised and less than 100℃,screen will show Fig.5.3-20 interface (the heating coil pre-heating stage). When temperature of heating coil(T3) rised and over than 100℃,screen will show Fig.5.3-21 interface (the first times pre-vacuum stage).
  • Page 35
    Fig.5.3-23 Fig.5.3-24 Fig.5.3-25 Fig.5.3-26 5.3.2.10.AFTER STERILIZATION EXHAUST STAGE(Fig.5.3-27) When the “sterilization time”countdown is over,the control system will open the exhaust-solenoid valve(Ev1) for exhausting untill pressure downs to P=0.000bar.LCD shows “00:01:29”time countdown. 5.3.2.11.DRYING STAGE(Fig.5.3-28) When the pressure inside the chamber downs to 0.000bar,the control system will open the vacuum pump-solenoid valve(Ev4) and evacuating and drying.LCD shows “00:08:59”time countdown.
  • Page 36
    WARNNING: If the result of sterilization is “FAILURE”, medical instruments of this time can not be used. And it’s need to contact professional engineer to check the machine. Fig.5.3-31 Fig.5.3-32 5.3.3 USER-DEFINNED PROGRAM Fig.5.3-33 Fig.5.3-34 Fig.5.3-35 Click “User-definned”(Fig.5.3-33) enter into interface Fig.5.3-34.Press “ ”to change the number of “Temperature/ Vacuum times/Sterilization time/Drying time”.
  • Page 37
    5.4 Test program When LCD shows main interface,Click “TEST”(Fig.5.4-1) enter into interface Fig.5.4-2. 5.4.1 the “vacuum test” program following steps: This test is performed in order to check the performance of the unit,particularly: -the efficiency of the vacuum pump; -the seal of the pneumatic circuit. Click “Vacuum Test”enter into interface Fig.5.4-3.Then press “…
  • Page 38
    In compliance with EN 13060:2004,the test requires a tightness test (less than or equal to 13mbar during the 10 minutes of test).If the leakage is more than this value,the outcome of the test is failed; the seal of the pneumatic circuit of the device must be chceked.
  • Page 39
    6.Maintenance Safety warnnings —Observe the intervals prescribed or shown in this manual.Activates menorandum messages to assist the users to performing Routine Maintenance Page 38 of 58…
  • Page 40
    Page 39 of 58…
  • Page 41
    Fig.6.2-1 Fig.6.2-2 Fig.6.2-3 Page 40 of 58…
  • Page 42
    Chapter 4.3.2 Chapter 4.3.2 Fig.6.2-4 Fig.6.2-5 Fig.6.2-6 Page 41 of 58…
  • Page 43
    6.2.1 Periodic Maintenance Page 42 of 58…
  • Page 44
    Extraordinary Maintenance Page 43 of 58…
  • Page 45
    solenoid valve.(Fig.6.3-1) The fuse on the internal card is of the type: 5X20-10A(AC 230V/50Hz/60Hz);5X20-20A(AC 110V/60Hz) 6.3.1 Rusting Fig.6.3-1 Page 44 of 58…
  • Page 46
    Scraping .1 Scraping Instructions 7.2 Resale A correct sterilization depends on the processes described below being carried out correctly; these are allequally important and care must be performed carefully. 1. Preparing the instrumnets to sterilize; 2. Packing; 3. Loading; 4. Sterilizing; 5.Preserving the sterilized instruments;…
  • Page 47
    The way how the sterilized instruments are packaged,and then sticked,determines the state of preservation of sterilization. The following material can be used as containers:metal containers with lids or perforated buttoms with filters in paper,pouches in paper Page 46 of 58…
  • Page 48
    Pouches with paper-polypropylene are excellent packageing systems of steam sterilizing for small sets of surgical instruments and The way how to load to sterilize is arranged is also important to the sterilization process condiderably .Always observe Page 47 of 58…
  • Page 49
    distance from others that they remain separate for the whole sterilization cycle. -Do not stack the trays on another and use the tray wupport.It is necessary to leave a space beteen each tray to allow for the circulation of steam during the sterilization phase and facilitate drying. When sterilizing the material in pouches, Fig.A3-1 Fig.A3-2…
  • Page 50
    It can perform three tests: The frequency of these tests should be performed following: This test is performed in order to check the performance of the unit,particularly: In compliance with EN 13060,the test requires a tightness test(less than or equal to 13mbar during the 10 minutes of test).if the leakage is more than this value,the outcome of the test is failed;the seal of the pneumatic circuit of the device must be checked.
  • Page 51
    Page 50 of 58…
  • Page 52
    ℃ ℃ ℃ 134 -04min-15min-3 134 -18min-15min-3 121 -20min-15min-3 Page 51 of 58…
  • Page 53
    Page 52 of 58…
  • Page 54
    APPENDIX 7: JN-23 Steam Sterilizer Hydraulic Circuit Safety vavle Opened-end Water and steam Condenser Solenoid Valve opend Solenoid Valve closed (one input and one output) separation device (Drain water) (one input and one output) Atmosphere Three-way piece Opend end(3#) Used-water tank (2#) Closed end Common end(1#)
  • Page 55
    The transformer output AC 21V Power Resume button Electronic valve closed(air filter, white, KD27 LED Light(N) The transformer Electronic valve closed(water pump, yellow, output AC 9V KD14 Electronic valve closed(drain water, brown, Electronic valve opened(vacuum pump, gray, Electronic lock DC12V output LED Light(L) KD16 Black…
  • Page 56
    APPENDIX 9:Alarm code and solutions of JN-23 Steam Sterilizer Alarm Code Release Alarm Alarm Cause Alarm Code Alarm Cause Solution “ ” If need abnormally exit during sterilization,you have to press button,then click “ ” “ ” “ ” ,LCD appear E00 alarm code and flashing.Press button again to relieve E 00…
  • Page 57
    Alarm Code Alarm Cause Solution Sterilization failed! (1)adjust the tightness of door seal may behave leakage.(2)Replace the door seal. E 04 (1)check the opened-solenoid valve(Ev1) and drain-solenoid valve(Ev5),take out and clean Pressure can not E 05 the s pool.(2)check three-way piece to used-water tank,open and clean it.(3)check the opened be exhausted! -Solenoid valve(Ev1) and drain-solenoid valve(Ev5) power is DC 24V.
  • Page 58
    Alarm Code Alarm Cause Solution Ω ≈ ℃ Temperature and Measure:temperature sensor resistance value( ) 1000+(Ambient temperature *3.8) E 10 Pressure do not (1)if measurement result is OK,check poor contact point between the sensor and the match! control board.(2)if measurement result is No,replac the temperature sensor. (1)check DC 24V d oor motor power supply indicator is lighted?
  • Page 59
    NOTE: 1.The sterilizer should be put on level worktable. 2.Be sure to use distilled water in order to prolong the use age of sterilizer. 3.Do not jam or cover the sterilizer radiator. 4.Sterilizer instrument should put in the instrument plate,each instrument should have some gap in order to make ventilation.

This manual is also suitable for:

Jn-23

Автоклавы Woson: коды ошибок и причины их появления.

Медицинский автоклав – прибор для стерилизации инструментов под повышенным давлением в закрытом пространстве.
Автоклавы Woson производятся в Китае с 2000-го года. Популярность на российском рынке автоклавы Woson обрели благодаря хорошему качеству продукции и сравнительно недорогой ценой.
Для обеспечения нормальной работы аппарата используется множество сложных узлов и агрегатов, которые, при несвоевременном обслуживании или использовании не дистиллированной воды, могут выходить из строя. В этом случае автоклав не закончит цикл стерилизации и будет выдавать ошибки. Практически на автоклава Woson коды ошибок имеют одинаковые значения. Появление ошибки на автоклаве говорит о том, что требуется вмешательство.
В этой статье будут описаны ошибки автоклавов Woson и возможные причины их возникновения. Прочитав данную статью Вам будет понятно, Вы в состоянии самостоятельно устранить неисправность или придется вызывать специалиста.

Ошибки автоклава Woson
1. E01 Медленный нагрев.
В большинстве случаев данная ошибка говорит о выходе из строя термодатчика парогенератора. Необходимо заменить этот датчик. Для этого нужно снять крышку автоклава. Парогенератор находиться под камерой стерилизации.
2. E02 Давление не может быть сброшено.
За сброс давления в камере автоклава отвечает клапан. Но перед ним находиться водный фильтр. Расположен он внутри камеры стерилизации. В инструкции к автоклаву указанно, как его выкрутить и почистить. Обычно эта процедура решает проблему, и если не помогло нужно осматривать систему дальше и ремонтировать клапан сброса давления.
3. E03 Открытая дверца во время цикла.
Автоклав «видит» открыта или закрыта дверца с помощью кнопки, она же датчик закрытия двери. Во время вакуумирования автоклав открыть вряд ли получиться. Открыть автоклав во время набора давления небезопасно, для этого на многих моделях стоит запорный механизм. При появлении данной ошибки нужно заменить датчик закрытия двери.
4. E04 Слишком низкое давление.
Для начала нужно проверить целостность уплотнительного кольца на дверки. Если она в порядке – отрегулировать дверцу. Как это сделать – описано в инструкции. Если это не помогает, смотрим в сторону парогенератора. Если он разборный – протянуть болты. На последних моделях Woson парогенератор не разбирается, в данном случае его нужно заменить.
5. E05 Слишком медленный спад давления.
Действия такие же, как при ошибке E02. Чистим водный фильтр, находящийся в камере стерилизации, если не помогло – разбираемся с клапаном, отвечающим за сброс давления.
6. E06 Медленное вакуумирование.
Медленное вакуумирование говорит о том, что есть утечки воздуха в автоклаве. Просмотреть уплотнительную резину двери. Отрегулировать дверцу согласно инструкции к автоклаву. Если ошибка не исчезла, скорее всего это связанно с разрывом шланга. Его требуется заменить.
7. E07 Переработка по времени.
Эта очень распространенная ошибка. Она говорит о том, что процесс автоклавирования не соответствует временным параметрам. Тут может быть одна из причин:
-Медленное вакуумирование.
-Медленный нагрев.
-Слишком медленный спад давления.
-Слишком медленный набор давления.
Для локализации неисправности нужно посмотреть, какой из процессов идет медленно, а за тем, действовать в соответствии E01, E05 или E06.

8. E08 Микропереключатель передней двери сломан.
Проверяется целостность провода микропереключателя, если все в порядке – меняется микропереключатель.
9. E09 Микропереключатель задней двери сломан.
То же, что и при ошибке E08.
10. E10 Низкий предварительный нагрев.
Эта ошибка связана с нагревом парогенератора. Для начала нужно проверить датчик температуры парогенератора. Затем проверить плотность его прилегания к парогенератору, сменить термопасту. Так же ошибка может быть связана с перегоревшим нагревательным элементом парогенератора. Нужно найти неисправный и заменить.
11. E11 Слишком высокая температура.
Данная ошибка возникает, когда нагревательный элемент стерилизационной камеры перегревает камеру. В большинстве случаев это связанно с выходом из строя «пробоем» симистора, управляющего нагревом. Снимается плата и симистор меняется.
12. E12 Слишком низкая температура.
Вышел из строя нагревательный элемент стерилизационной камеры. Его нужно проверить. Если он целый, то смотрим цепь питания, скорее всего будет сгоревшим симистором. Тогда делаем, как в E11.
13. E13 Автоматическая подача воды нарушена.
За подачу воды в автоклаве отвечает помпа. Она соединена трубками с резервуаром и парогенератором. Для начала нужно проверить целостность трубок и соединений. Если все в порядке – проблема с помпой. После ее замены на аналогичную неисправность пропадет.
14. E17 Качество воды в резервуаре суперскалярное.
В современных автоклавах устанавливают датчик качества воды. Если была залита вода не соответствующего качества, то автоклав выдаст такую ошибку. Потребуется слить старую и залить новую воду соответствующего качества. Если не поможет, рекомендуется проверить дистиллятор. Еще такая ошибка может появится, если датчик качества воды загрязнен. Он выглядит, как два железных стержня и находится в баке с чистой водой. Его нужно почистить.
15. E23 Давление и температура не совпадают.
Когда в стерилизационной камере температура вышла на заданный режим, а давление не вышло на нужный параметр, возникает данная ошибка. Ее вызывает забитый парогенератор, что вызвано использованием некачественной воды. Нужно промыть парогенератор раствором лимонной кислоты, 100 грамм кислоты на литр воды. Слить воду из резервуара и залить с лимонной кислотой. Провести 3-4 цикла автоклавирования. Если не помогло, придется разбирать и чистить парогенератор. Если он не разборный, поможет только замена.
16. E24 Отсутствует стерильная вода.
При появлении этой ошибки, делается тоже, что и при ошибке E17.
17. E25 Загрязнен воздушный фильтр.
В данном случае действия простые – открывается инструкция. В ней указан тип фильтра и место, где он находиться. Так же в инструкции описано, как его заменить. Для этого не потребуется инструмент и технические знания.
18. E31, E32, E33, E34 Температура Сенсора 1, 2, 3, 4 Соединение отсутствует.
В автоклаве есть несколько датчиков температуры:
-Датчик температуры парогенератора,
-Датчик внутренней температуры стерилизационной камеры.
-Датчик внешней температуры стерилизационной камеры.
-Датчик температуры двери.
Соответствие номера и датчика указаны в документах. Замена датчика температуры решает данную неисправность.
19. E51, E52, E53, E54 Датчик температуры высокий уровень выхода.
Данная ошибка может говорить о неисправности микроконтроллера, управляющим всеми процессами автоклава. Данную поломку починит не так просто, потребуется обращение в инженерный отдел производителя или замена платы полностью.
20. E71 Сенсор давления низкий уровень выхода.
Данная ошибка говорит о выходе из строя датчика давления. Если замена не решает проблему, все указывает на ошибку микроконтроллера автоклава.
21. E72 Сенсор давления высокий уровень выхода.
Так же, как и при ошибке E71.

На разных моделях номера ошибок могут отличаться. Их расшифровки всегда указаны в документации с автоклавом.
В данной статье указаны основные методы ремонта автоклава, если стать помогла решить Вам проблему – отлично. Если же Вы не можете своими силами осуществить ремонт, обращайтесь! Наши квалифицированные специалисты быстро и качественно устранят неисправность!

Ошибка указывается на дисплее надписью «Error», номером ошибки и ее кратким обозначением.

Сообщения об ошибках могут появляться до начала выполнения программы (при включении питания или позже) и во время работы программы.

Если об ошибке сообщено во время работы программы, программа будет приостановлена. Этому может сопутствовать нормализация давления в автоклаве, и в этом случае появятся надписи «Pressure release» (снятие давления), или «Ventilation» (вентиляция), и «End» (конец).

После прерывания дисплей будет переменно указывать сообщение об ошибке и «Quit with «-» кнопка» (выход кнопкой + кнопка), затем появится надпись «Terminate End » (остановка программы). Нажатие «-» удаляет сообщение об ошибке (если функция не является продолжительной) Дверцу автоклава нельзя открыть до тех пор, пока Вы не избавитесь от сообщения об ошибке. Если программа была прервана таким образом, груз в автоклаве следует считать нестерильным. Мы рекомендуем Вам разгрузить автоклав, провести стерилизацию с пустой камерой (сушку можно исключить), а затем перезагрузить автоклав и повторить прерванный цикл.
Если подключен внешний принтер и выбран режим «Immed. print-out? Yes» («Непосредственная распечатка? Да»), по прерыванию программы будет автоматически распечатана запись об этом.

Распечатка указывает полное название ошибки, и, в случае прерывания программы, дается надпись «Load not sterile» («Груз нестерилен»). В данном перечне указаны сообщения об ошибках, их причины и возможные действия по их устранению:

Сообщение об ошибке Причина/Рекомендуемые действия
Превышено время достижения разряжения, стравливания давления, или давление сушки:

  • Проверьте уплотнительные прокладки на дверце. Чтобы края крышки в камеру были не повреждены и находились в чистом состоянии.
  • Проверьте, чтобы автоклав был установлен правильно (см. параграф 2.3).
  • Проверьте, чтобы стоку конденсата не препятствовали свалившиеся инструменты, обрывки фильтровальной бумаги и др. на дне вакуумной камеры.
  • Проверьте, чтобы фильтр не был заблокирован (с передней стороны основания вакуумной камеры).
  • Проверьте утечки, с помощью программы «Vacuum test».

Если это происходит постоянно, проинформируйте Вашего дилера-специалиста.

Превышено время нагрева в процессе разряжения воздуха или достижения требуемого давления для стерилизации:

  • Возможные причины данной неисправности могут быть следующими:
  • Превышена максимально допустимая загрузка камеры.
  • Слишком низкое напряжение сети; проверьте сетевое питание; попробуйте подключить автоклав к другой электрической цепи.
  • Потеря воды в следствии утечек, различных образований или скопления воды.
  • Не допускайте скопления воды в различных углублениях инструментов во время стерилизации: переворачивайте чаши, стаканы верхней стороной (вогнутостью) вниз.
  • Внимание: не допускается производить стерилизацию без использования лотка.
  • После устранения приведенных выше неисправностей, нажмите кнопку Reset на приборе.
  • Внимательно: После нажатия на Reset, выполните холостую стерилизацию Fast program (стерилизация с пустой камерой)

Если это происходит постоянно, проинформируйте Вашего дилера-специалиста.

Превышено время нормализации давления:

  • Убедитесь, что выпускные клапаны на задней панели не заблокированы
  • Если это происходит постоянно, проинформируйте Вашего дилера-специалиста.
Длительность программы превышает возможность внутренних часов:

Если это происходит постоянно, проинформируйте Вашего дилера-специалиста.

Неплотно закрыта дверца:

  • Нажимайте ручку вниз, пока не защелкнется замок (дисплей должен показать «Door closed» – «Дверца закрыта»)!
  • Если это происходит постоянно, проинформируйте Вашего дилера-специалиста.
В момент запуска программы открыт регулятор уровня (сообщение об ошибке сразу после запуска) или превышено контрольное время регенерации воды в период работы:

  • Эта проблема может возникнуть в интенсивном режиме работы. Подождите две минуты и повторите запуск.

Если это происходит постоянно, то проинформируйте своего дилера-специалиста.

Максимально возможное время запирания дверцы:

Убедитесь, что засов замка двигается свободно.

  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Во время работы не закрывайте контроллер подачи воды (см. сообщение «Warning no feed water» – страница 34).
Внутреннее тестирование датчиков температуры, давления и проводимости показывает недопустимое отклонение. Ошибка заявляется при выполнении и в период работы:

  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Прогрев не обеспечил необходимой температуры в контрольный период времени:

Если это происходит постоянно, установите режим «Automatic preheating No» ( «Предварительный нагрев Нет») и проинформируйте Вашего дилера(см. раздел6.5).

Превышена максимальная температура прогрева:

  • Если это происходит постоянно, отмените прогрев и проинформируйте своего дилера.

Malfunction 23:

Current

Превышено контрольное время нормализации давления:

  • Убедитесь, что в камере под дверцей не заблокирован фильтр.
  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Превышена допустимая погрешность преобразования внутренних сигналов:

  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Превышена допустимая разница температур между двумя температурными датчиками:

Если это происходит постоянно, то проинформируйте своего дилера-специалиста.

Противоречивость или утеря данных в модуле памяти. Это может быть результатом помех электропитания или недостаточного напряжения аккумулятора. При сбрасывании часы автоматически установят 00.00 и счетчику загрузок будет присвоено значение из EEPROM. Все записи в памяти будут стерты:

  • После отмены сообщения об ошибке, переустановите дату и время (см. раздел 6.4) и перегрузите.
  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
После старта программы прервана подача питания. Сообщение об ошибке появляется после восстановления питания:

  • Проверьте внешнюю электросеть, если ошибки не найдены, проинформируйте обслуживающий персонал.

Если утеря энергии произошла в момент, когда камера была под давлением, появится дополнительное напоминание простерилизовать стерильный фильтр, так как он мог потерять стерильность:

  • Достаньте стерильный фильтр в задней панели автоклава.
  • Простерилизуйте фильтр в ускоренном режиме.
  • Поставьте фильтр на место.
Превышено время нагнетания давления:

  1. Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Не достигнута минимальная температура стерилизации:

  • Сократите объем загрузки.
  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Превышена максимальная температура стерилизации:

  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Давление падает ниже допустимого уровня:

  • Сократите объем загрузки.
  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Превышено максимально допустимое давление:

  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Большая разница между измеренной и теоретически ожидаемой температурой:

  1. Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Превышено контрольное время нормализации давления на этапе сушки:

  • Убедитесь, что в камере под дверцей не заблокирован фильтр.
  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Превышено контрольное время нагнетания давления на этапе сушки:

  • Убедитесь, что стерильный фильтр не засорился, при необходимости замените.
  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.
Превышено время создания вакуума на этапе сушки:

  • Убедитесь, что фильтр позади крышки не засорился и не был заблокирован.
  • Если это происходит постоянно, то проинформируйте своего дилера-специалиста.

  • Автокад часто выдает фатальную ошибку
  • Автокад фатальная ошибка при штриховке
  • Автокад фатальная ошибка при сохранении
  • Автокад фатальная ошибка неверное имя файла настройки или путь
  • Автокад фатальная ошибка unhandled e06d7363h exception